Also not covered: PCR enclosures (HEPA-filtered sterile environments for PCR experiments) and powder containment enclosures (approved with HEPA filters for weighing toxic compounds, etc.) are
not serviced through the Primary Barrier Program. Service for these units must be completed according to manufacturers' recommendations by a laboratory-procured contractor/vendor.
Note on airflow issues: Chemical fume hoods, LEV units, and Class II Type B ducted BSCs are connected to building exhaust ventilation systems. Airflow problems traced to building ventilation must be resolved by ORF maintenance personnel before a unit can be recertified. Minor repairs (e.g., light replacement) may be completed by ORF; major repairs are obtained by the laboratory through a third-party vendor.
Equipment Configuration Restrictions
Certain biological safety cabinet configurations are not permitted, including:
- Class II Type C cabinets (thimble/canopy connection)
- Cabinets fitted with ultraviolet (UV) disinfection lights.
Contact the Primary Barrier Program before purchasing new equipment to confirm it meets current requirements.