Certification, repair, and decontamination services for primary barrier equipment including biological safety cabinets, chemical fume hoods, and local exhaust ventilation units.

📞 PrimaryBarrier@mail.nih.gov | (301) 496-3457

🔧​

Certification & Repair Process

​What happens after you submit a request, and how airflow issues are resolved.

Routine Certification

1 Lab or facility personnel submit a service request through CERTS.
2 A certified technician performs field testing — verifying airflow, containment, and filter integrity as applicable to the equipment type.
3 If the equipment passes, a certification sticker/label and documentation are issued.
4 If the equipment fails, it is taken out of service until repairs are completed and it is recertified.

Resolving Airflow Issues (Fume Hoods, LEVs, Ducted BSCs)

1 Airflow issues connected to building ventilation are reported to ORF for resolution.
2 ORF adjusts building airflow; minor repairs (e.g., light replacement) may also be completed by ORF.
3 Once airflow has been adjusted, lab or facility personnel submit a primary barrier service request for recertification following the building adjustment.
4 Major repairs are obtained by the laboratory through a third-party vendor.
Repairs

​Repair Requests

Submit a repair request through CERTS as soon as equipment is identified as non-functional. Equipment should not be used for hazardous work until it has been repaired and, if applicable, recertified.

Decontamination

​Decontamination Requests

Decontamination is required before a biological safety cabinet is moved, serviced, repaired, or disposed of. Submit a decontamination request through CERTS in advance of any planned relocation or major service work.